Bench Top Freeze Dryer

The freeze-dryer has been exclusively designed for the freeze-drying of solid or liquid products in ampoules, vials or dishes. The freeze-dryer is suitable for freeze-drying solid substances and aqueous solutions (e.g. bacteria and virus cultures, blood plasma, serum fractions, antibodies, sera, vaccines and pharmaceutical products such as chloramphenicol, streptomycin, vitamins, ferments and plant extracts for biochemical tests). The freeze dryer is mainly used in industries such as biomedicine, chemical industry, food and environmental testing. 

ModelAFD-10SAFD-10TAFD-10PAFD-10PT
TypeStandardStopper8 port manifold8 port manifold
and stopper
Ice condenser temp temperature-55°C
Ice condenser Capacity3Kg/24h
Freeze drying area 0.12m20.09m2
Drying Vacuum≤10Pa
Freezing temperature pull down time 30 minutes  (from 32°C to -55°C)
Freeze drying time24 hours
Controller 7" PLC touch screen with  microprocessor 
Sample temperature displayYes, display 0.1°C accuracy
Ice condenser temperature displayYes display 0.1°C accuracy
Data RecordSample temperature/Prefrozen temperature
Vacuum displayYes,1pa
Histroical record display Yes
USB Download functionYes
Votlage/Current220~240V/50HZ/9A
Main unit dimension582*541*374
Overeal (W*D*H,mm)582*541*824    582*541*884 582*541*824 582*541*884
Exterior Material Powder coated steel, table is SUS304
Condenser trap MaterialSUS304
Eggplant bottle quantity/8/8
Shelf layer   4 layers       3 layers   4 layers   3 layers
Material tray dimensionф200mm*H20mm
Material tray capacity (based on 10mm thickness)300
Total material load capacity(ml)12009001200900
Condener trap φxH(mm)φ215mm×H193.5mm
Adjustbale feetYes
Shelf heaterYes
Heater defrostNo
LN2 filling port No
-80°C condenser temperatureNo


Request a Quote

If you have any questions or comments, please fill out the form